DIGITAL RADIOGRAPHY SYSTEM

Ceiling Suspended Tube Stand

The Ceiling mounted tube provides

tracking with the upright stand and

table detectors improving workflow,

throughput and efficiency.

Selectable Two Style

FDR Smart X is Fujifilm’s newly

developed X-ray system, providing

multi-function, high-quality,

cost-effective X-ray solutions.

The Smart X offers ceiling suspended

X-ray tube configurations as well as a

floor mounted X-ray tube option for use

with the upright stand and table,

providing flexible solutions for your

imaging department.

Touch Screen Console

Compatible with Fujifilm’s latest D-EVO II

series DR panels with their advanced

features and high image quality.

Also CR cassette and F/S cassette are

available for FDR Smart X.

Compatible and Optimal X-ray Solution

A Digital display on the tube head

provides easy access to

exam parameters

in the X-ray room.

The Floor mounted tube option,

provides flexibility to install in limited

space, maximizing the imaging

systems available in the department.

02

Floor Mounted Tube Stand

03

Wall Bucky Stand Automatic Tilting

Automatic and manual tilting of the upright bucky stand is available*.

Providing an improved workflow for the radiographers and an improved

experience to those patients in wheelchairs.

*Different stand is lined-up for each

Multi function stand and table supporting efficient workflow

Elevating Patient Table

Control Panel of

Automatic Tilting Wall Stand

The ability to store preset positions and move the wall

stand automatically to those positions provides a quick and

efficient workflow.

A motorized height adjustable floating top table with a weight limit of

300 kg. Synchronisation (Tracking) of the tube with both upright and

table detectors. Available with ceiling and floor mounted systems.

Remote Control

• Auto Positioning (User Programmable)

• Control of Synchronization with Tube Stand

• Motorized Collimation & Lamp Control

• Motorized Tilting Movement

• Motorized Vertical Movement

Dual Reference Rotation

Removable Grid

Removable grids in both the table

and upright stand, allow flexible

working (such as Pediatrics).

Rotating Tray

Rotating tray is also available for

table. Direction of the panel can

be changed according to the

patient position.

Not only top reference, but also center rotation can be selected when using 14”

17”

DR panels. This provides flexible solutions, enabling exposure of areas other than the chest

without taking out the panel and insert again in rotated direction or using larger panel.

Automatic Tracking

Tube head and bucky move synchronized with each other supporting positioning for exposure.

Automatic Connection

The SE cable + connector are already built in the tray, thus, just setting

the DR panels into the tray, DR panels are automatically connected.

Also, there is no need to handle the cables around the bucky.

Combined with rotation function, this will greatly improve efficiency

and remove the stress of the workflow.

Automatic

Tracking

Manual

Operation

Manual

Operation

Manual

Operation

Manual

Automatic

Operation

Tracking

Automatic

Tracking

Automatic

Tracking

Fail Safe Switch

Fail safe switch detects panel in side the tray / bucky. When the panel is not inside the bucky or not inserted properly,

exposing X-ray is unable. Thus, false exposure can be prevented.

Motorized Vertical

Synchronization with wall stand

Motorized Vertical

Synchronization with table

The table bucky tracks after the

horizontal movement of the

tube head (same as floor

mounted tube head)

The table bucky tracks after the

rotating motion of the tube

head (same as floor mounted

tube head)

04

05

Integrated and optimized workflow and High Image Quality

Console

Technique select buttons

*From Console

*From X-ray operation

Connected modalities (panels / cassettes)

are displayed both in Console and X-ray

controller. By simply selecting the button,

the modality can be changed and they are

synchronized between Console and X-ray

controller.

Status display for DR panel

When DR panel is used, it is available to

confirm its status; wireless connection

level, battery charge level, etc.

X-ray operation

Integrated Console

Realizing optimized workflow

Detector Console and X-ray controller are integrated,

displaying in the same monitor. Setting the exposure

condition and selecting the exposure menu can be done

in the same PC.

High definition, made smarter.

FDR D-EVO II

Improved image quality through Fujifilm’s innovative

designs like ISS and noise reduction circuit allows

smarter imaging.

17

17 inch model14

17 inch model

Compatible for DR and CR systems

Both DR panel and CR cassette can be used for FDR Smart X.

(Also, F/S cassette is available as mentioned previously) They

can be used in the same system at the same time, either the

workflow is the same, and the image appears in the console.

(Using F/S cassette, image does not appear)

FDR Smart XIntegrated ConsoleFCR Reader

ISS system reading technology promotes

high sensitivity

Equipped with an indirect conversion system flat detector display

using ISS, which bonds optical sensors (TFT) to the X-ray

irradiation side unlike traditional flat detector displays. This

greatly suppresses scattering and attenuation of X-ray signals,

creating sharp images with low doses of X-rays.

Fujifilm noise reduction circuit improves

detector sensitivity in high absorption regions

The uniquely developed noise reduction circuit reduces noise

in the image. It achieves 1.7 times the DQE of existing

systems with a 0.03 mR dose. In particular, granularity of

low-concentration regions such as the heart and mediastinum

is dramatically improved.

Dynamic Visualization ll

Optimizing Contrast and Density with latest Exposure Data Recognizer

X-ray

X-ray

TFT detector

D

Q

E

Optical signal image

reaching TFT detector

Optical signal image

reaching TFT detector

Approx.

170

%

up

FDR D-EVO

FDR D-EVO II

Contrast and density can automati-

cally adjusted through recognizing

each body part and object based on

the estimated 3D information.

Conventional

06

Scintillation

layer

TFT detector

Scintillation

layer

0.010.030.11

With additional major increases in sensitivity in

low-concentration regions (heart, mediastinum)

Conventional method

Dynamic Visualization IIConventional Dynamic Visualization II

ISS system reading technology

07

Medical CenterCountry side

*Unstable electricity supply

Hospital

Capacitor Assisted Generator

Perfect for the medium sized X-ray

department. No power supply

modification necessary 40kW Single

Phase Input power 3kVA

UPS Generator

Supplies power for 5 hours (or approximately

500 shots) without electricity. The ideal

solution for regions with unstable power

supply 40kW Single Phase Input power 1kVA

Line Powered Generator

We can adjust our generator to

the workload required in a big hospital.

40kW/52kW/68kW/82kW Three phase.

This is a high frequency X-ray generator

FDR Smart FGXR series

FGXR-32S

Generator Model

Output Rating

230VAC,

1

Φ

, ±10%

GXR-32

FGXR-C32S

GXR-C32

32kW

110-120/

220-230VAC,

1

Φ

, 3KVA

FGXR-U32S

GXR-U32

FGXR-40S

GXR-40

FGXR-C40S

GXR-C40

40kW

FGXR-U40S

GXR-U40

FGXR-52S

GXR-52

FGXR-C52S

GXR-C52

FGXR-68S

GXR-68

68kW

FGXR-82S

GXR-82

82kW52kW

100-240VAC,

1

Φ

,

1KVA

400/480VAC,

3

Φ

,

±10%

110-120/

220-230VAC,

1

Φ

, 3KVA

100-240VAC,

1

Φ

, 1KVA

230VAC,

1

Φ

, ±10%

400/480VAC,

3

Φ

,

±10% (option)

400/480VAC,

110-120/

220-230VAC,

1

Φ

, 3KVA

400/480VAC, 3

Φ

, ±10%

DIGITAL RADIOGRAPHY SYSTEM

3

Φ

,

±10% (option)

Line Frequency

kV Range

40~125kV, 1kV step

(150kV option)

50/60Hz (Outside North America)

40~150kV,

1kV step

40~125kV,

1kV step

(150kV option)

40~150kV, 1kV step

mA Range

Timer Range

mAs Range

10 to 400mA10 to 500mA

0.001 to 10 sec, 38 steps

0.1 to 500mAs (Optional higher mAs)

10 to 640mA

10 to 800mA10 to 1000mA

Max. Power Output

400mA@80kV

320mA@100kV

250mA@125kV

500mA@80kV

400mA@100kV

320mA@125kV

640mA@81kV640mA@81kV

500mA@104kV500mA@104kV

400mA@130kV400mA@130kV

320mA@150kV

800mA@85kV

640mA@106kV

500mA@136kV

400mA@150kV

1000mA@82kV

800mA@102kV

640mA@128kV

500mA@150kV

Specifications are subject to change without notice.

All brand names or trademarks are the property of their respective owners.

All products require the regulatory approval of the importing country.

For details on their availability, contact our local representative.

Ref. No. XB-1044E

(SK·17·01·F1079·F9711) Printed in Japan ©2017 FUJIFILM Corporation


发布者:admin,转转请注明出处:http://www.yc00.com/num/1712729150a2111742.html

相关推荐

发表回复

评论列表(0条)

  • 暂无评论

联系我们

400-800-8888

在线咨询: QQ交谈

邮件:admin@example.com

工作时间:周一至周五,9:30-18:30,节假日休息

关注微信