Essential Principles Checklist
Identity of IVD medical device:
Blood glucose monitoring system(URIT-25、URIT-26、URIT-28)
Essential Principle Applicable to
the device?
A General Requirements
A1 Medical devices should be designed and
manufactured in such a way that, when used under
the conditions and for the purposes intended and,
where applicable, by virtue of the technical
knowledge, experience, education or training, and the
medical and physical conditions of intended users,
they will perform as intended by the manufacturer
and not compromise the clinical condition or the
safety of patients, or the safety and health of users or,
where applicable, other persons, provided that any
risks which may be associated with their use
YES
Method Method Reference Reference to
Used to Supporting Controlled
Demonstrate Documents
Conformity
Recognised
EN ISO 14971:2012 CE-019
standard
EN 61010-2-101:2002
EN ISO 15197:2003
CE-013
CE-021
1 constitute acceptable risks when weighed against the
benefits to the patient and are compatible with a high
level of protection of health and safety.
A2 The solutions adopted by the manufacturer for the
design and manufacture of the devices should
conform to safety principles, taking account of the
generally acknowledged state of the art. When risk
reduction is required, the manufacturer should
control the risks so that the residual risk associated
with each hazard is judged acceptable. The
manufacturer should apply the following principles
in the priority order listed:
• identify known or foreseeable hazards and estimate
the associated risks arising from the intended use
and foreseeable misuse;
YES
Recognised
standard
EN ISO 14971:2012
EN 61010-2-101:2002
EN ISO 15197:2003
EN 62366:2008
EN ISO 18113-4:2011
EN ISO 18113-5:2011
CE-019-1,CE-019-2,CE-019-3
CE-013
CE-021
CE-017
CE-011
MANUAL
(URIT G25 / URIT G26 /
URIT G28)
(URIT-25/URIT-26/URIT-28)
OPERATING
• eliminate risks as far as reasonably
practicable through inherently safe design and
manufacture;
reduce as far as reasonably practicable the
remaining risks by taking adequate protection
measures, including alarms; and
inform users of any residual risks.
YES
•
•
A3 Medical devices should achieve the performance
Recognised
EN ISO 15197:2003 CE-021
2 intended by the manufacturer and be designed and
manufactured in such a way that, during normal
conditions of use, they are suitable for their intended
purpose.
A4 The
characteristics and performances referred to in Clauses
YES
A1, A2 and A3 should not be adversely affected to such a degree
that the health or safety of the patient or the user and, where
applicable, of other persons are compromised during the lifetime of
the device, as indicated by the manufacturer, when the device is
subjected to the stresses which can occur during normal conditions
of use and has been properly maintained in accordance with the
manufacturer’s instructions.
standard
EN 62366:2008
EN 62304:2006
CE-017
CE-016
Recognised
standard
EN13612:2002
CE-022
A5
Medical devices should be designed, manufactured and
YES
packaged in such a way that their characteristics and performances
during their intended use will not be adversely affected by transport
and storage conditions (for example, fluctuations of temperature
and humidity) taking account of the instructions and information
provided by the manufacturer.
Recognised
standard
EN ISO 15197:2003
CE-021
A6 All known and foreseeable risks, and any undesirable effects,
YES
should be minimised and be acceptable when weighed against the
benefits of the intended performance of medical devices during
normal conditions of use.
Recognised
standard
EN ISO 14971:2012 CE-019
B Essential Principles applicable to Medical Devices
/
other than IVD Medical Devices
C Essential Principles applicable to IVD Medical
/
/
/
3 Devices
C1 Chemical, physical and biological properties
C1.1 The IVD medical devices should be designed and manufactured in such a
way as to ensure the characteristics and performance referred to in Section 6.
Particular attention should be paid to the possibility of impairment of analytical
performance due to incompatibility between the materials used and the
specimens and/or analyte (measurand) to be detected (such as biological
tissues, cells, body fluids and micro-organisms), taking account of its intended
purpose.
C1.2 The IVD medical devices should be designed, manufactured and
packaged in such a way as to minimize the risk posed by contaminants and
residues to the persons involved in the transport, storage and use of the devices
and to patients, taking account of the intended purpose of the device.
C1.3 The IVD medical devices should be designed and manufactured in such a
way as to reduce as far as reasonably practicable and appropriate the risks
posed by substances that may leach or leak from the IVD medical device.
Special attention should be given to substances which are carcinogenic,
mutagenic or toxic to reproduction.
YES
EN ISO 14971:2012
CE-019
Recognised
standard
YES
Recognised
standard
EN ISO 14971:2012 CE-019
NO. This device
does not have
the transudates
or leakages
which are
carcinogenic,
mutagenic or
toxic to
reproduction.
4 C1.4 IVD medical devices should be designed and manufactured in such a way
YES
as to reduce as far as reasonably practicable and appropriate risks posed by the
unintentional ingress or egress of substances into or from the IVD medical
device taking into account the device and the nature of the environment in
which it is intended to be used.
Recognised
standard
EN ISO 14971:2012 CE-019
C2 Infection and microbial contamination
C2.1 The IVD medical devices and manufacturing
processes should be designed in such a way as to
eliminate or to reduce as far as reasonably
practicable and appropriate the risk of infection to
user, professional or lay, or, where applicable,
other person . The design should:
• allow easy and safe handling;
YES
EN ISO 14971:2012
CE-019
Recognised
standard
and, where necessary:
• reduce as far as reasonably practicable and
appropriate any microbial leakage from the
IVD medical device and/or microbial exposure
during use; and
prevent microbial contamination of the IVD
medical device or specimen where applicable,
by the user, professional or lay, or other person.
NO This
does
device
not
•
C2.2 IVD medical devices labelled either as sterile or as having a special
microbiological state should be designed, manufactured and packaged to
5 ensure they remain so when placed on the market and remain so under the
transport and storage conditions specified by the manufacturer, until the
protective packaging is damaged or opened.
C2.3 IVD medical devices labelled either as sterile or as having a special
microbiological state should have been processed, manufactured and, if
applicable, sterilized by appropriate, validated methods.
C2.4 IVD medical devices intended to be sterilized should be manufactured in
appropriately controlled (e.g. environmental) conditions.
contain
microbes.
NO This device
does not
contain
microbes.
This device
does not
contain
microbes.
This device
does not
contain
microbes.
This device
does not
contain
biological
source
materials
NO
C2.5 Packaging systems for non sterile IVD medical devices should maintain
the integrity and cleanliness of the device.
NO
C3 IVD medical devices incorporating materials of biological origin
C3.1 Where IVD medical devices include tissues, cells and substances
originating from animals, the processing, preservation, testing and handling of
tissues, cells and substances of animal origin should be carried out so as to
provide optimal safety for user, professional or lay, or other person.
In particular, safety with regard to viruses and other transmissible agents
should be addressed by implementation of validated methods of elimination or
inactivation in the course of the manufacturing process. This may not apply
to certain IVD medical devices if the activity of the virus and other
transmissible agent are integral to the intended purpose of the IVD medical
device or when such elimination or inactivation process would compromise the
NO
6 performance of the IVD medical device.
National regulations may require that the manufacturer and/or the Regulatory
Authority retain information on the geographical origin of the animals.
C3.2 Where IVD medical devices include human tissues, cells and substances,
the selection of sources, donors and/or substances of human origin, the
processing, preservation, testing and handling of tissues, cells and substances
of such origin should be carried out so as to provide optimal safety for user,
professional or lay, or other person.
In particular, safety with regard to viruses and other transmissible agents
should be addressed by implementation of validated methods of elimination or
inactivation in the course of the manufacturing process. This may not apply
to certain IVD medical devices if the activity of the virus and other
transmissible agent are integral to the intended purpose of the IVD medical
device or when such elimination or inactivation process would compromise the
performance of the IVD medical device.
C3.3 Where IVD medical devices include cells and substances of microbial
origin, the processing, preservation, testing and handling of cells and
substances should be carried out so as to provide optimal safety for user,
professional or lay, or other person.
In particular, safety with regard to viruses and other transmissible agents
should be addressed by implementation of validated methods of elimination or
inactivation in the course of the manufacturing process. This may not apply
to certain IVD medical devices if the activity of the virus and other
transmissible agent are integral to the intended purpose of the IVD medical
device or when such elimination or inactivation process would compromise the
performance of the IVD medical device.
NO This device
does not
contain
biological
source
materials
NO This device
does not
contain
biological
source
materials
7 C4 Environmental properties
C4.1 If the IVD medical device is intended for use in combination with other
devices or equipment, the whole combination, including the connection system
should not impair the specified performance of the devices. Any restrictions
on use applying to such combinations should be indicated on the label and/or in
the instructions for use.
YES
EN ISO 14971:2012
CE-019
Recognised
standard
C4.2 IVD medical devices should be designed and manufactured in such a way
as to remove or reduce as far as reasonably practicable and appropriate:
C4.2.1
C4.2.2
C4.2.3
the risk of injury to user, professional or lay, or other person in
the risk of use error due to the ergonomic features, human factors
risks connected with reasonably foreseeable external influences or
EN ISO 14971:2012
CE-019
YES
YES
YES
connection with their physical and ergonomic features;
and the environment in which the IVD medical device is intended to be used;
environmental conditions, such as magnetic fields, external electrical and
electromagnetic effects, electrostatic discharge, pressure, humidity, temperature
or variations thereof;
C4.2.4 the risks associated with the use of the IVD medical device when it
Recognised
standard
Recognised
standard
Recognised
standard
EN ISO 14971:2012
EN ISO 14971:2012
CE-019
CE-019
CE-012
CE-013
CE-019
EN 61326-2-6:2006
EN 61010-2-101:2002
EN ISO 14971:2012
YES
comes into contact with materials, liquids, and gases to which it is exposed
during normal conditions of use;
C4.2.5 the risk associated with the possible negative interaction between
software and the environment within which it operates and interacts;
C4.2.6 the risks of accidental penetration of substances into the IVD medical
device;
Recognised
standard
Recognised
standard
Recognised
standard
YES
EN ISO 14971:2012 CE-019
YES
EN ISO 14971:2012 CE-019
8 C4.2.7 the risk of incorrect identification of specimens/samples;
C4.2.8 the risks of reasonably foreseeable interference with other devices such
as carry over between IVD medical devices.
C4.3 IVD medical devices should be designed and manufactured in such a
way as to minimize the risks of fire or explosion during normal use and in
single fault condition. Particular attention should be paid to IVD medical
devices whose intended use includes exposure to or use in association with
flammable substances or substances which could cause combustion.
C4.4 IVD medical devices should be designed and manufactured in such a
way that adjustment, calibration, and maintenance, where such is necessary to
achieve the performances intended, can be done safely.
C4.5 IVD medical devices should be designed and manufactured in such a
way as to facilitate the safe disposal of any waste substances.
C5 Performance characteristics
YES
YES
Recognised
standard
Recognised
standard
Recognised
standard
EN ISO 14971:2012 CE-019
EN ISO 14971:2012 CE-019
YES
EN 61010-2-101:2002 CE-013
YES
Recognised
standard
EN 61010-2-101:2002 CE-013
YES
Recognised
standard
EN ISO 14971:2012 CE-019
EN ISO 15197:2003
EN13612:2002
C5.1 IVD medical devices should be designed and
manufactured in such a way that the performance
characteristics support the intended use, based on
appropriate scientific and technical methods. In
particular, where appropriate, the design should
address sensitivity, specificity, accuracy which is
trueness and precision (repeatability and
YES
Recognised
standard
CE-021
CE-014
9 reproducibility), control of known relevant
interference and limits of detection.
These performance characteristics need to be maintained
during the lifetime of the IVD medical device as indicated
by the manufacturer.
C5.2 Where the performance of devices depends on the
use of calibrators and/or control materials, the
traceability of values assigned to such calibrators
and/or control materials should be assured through
available reference measurement procedures and/or
available reference materials of a higher order.
C5.3 Wherever possible values expressed numerically
should be in commonly accepted, standardised
units, and understood by the users of the device.
C6 Protection against radiation
C6.1 IVD medical devices should be designed,
manufactured and packaged in such a way that
exposure of user, professional or lay, or other
person to the emitted radiation (intended,
unintended, stray or scattered) is reduced as far as
reasonably practicable and appropriate.
C6.2 When IVD medical devices are intended to emit
YES
Recognised
standard
EN ISO 15197:2003
CE-006
YES
Recognised
standard
EN ISO 15197:2003
CE-021
CE-012
YES
Recognised
standard
EN 61326-2-6:2006
NO。 This device does
10 potentially hazardous, visible and/or invisible
radiation, they should as far as reasonably
practicable and appropriate be:
• designed and manufactured in such a way
as to ensure that the characteristics and
the quantity of radiation emitted can be
controlled and/or adjusted; and
fitted with visual displays and/or audible
warnings of such emissions.
not intend to emit
radiation.
•
C7 IVD medical devices that incorporate software
and standalone IVD medical device software
C7.1 For IVD medical devices which incorporate
software or for standalone software that are IVD
medical devices in themselves, the software must
be validated according to the state of the art taking
into account the principles of development
lifecycle, risk management, verification and
validation.
C8 IVD medical devices connected to, or equipped
with, an energy source
C8.1 IVD medical devices where the safety of the patient
depends on an internal power supply in the IVD
YES
Recognised
standard
EN ISO 14971:2012
EN 62304:2006
CE-019
CE-016
YES
Recognised
standard
EN ISO 14971:2012
CE-019
CE-021
11 medical device, should be equipped with a means
of determining the state of the power supply.
C8.2 IVD medical devices should be designed and
manufactured in such a way as to reduce as far as
reasonably practicable and appropriate the risks of
creating electromagnetic interference which could
impair the operation of this or other devices or
equipment in the usual environment.
C8.3 IVD medical devices should be designed and
manufactured in such a way as to provide an
adequate level of intrinsic immunity to
electromagnetic disturbance to enable them to
operate as intended.
C8.4 IVD medical devices should be designed and
manufactured in such a way as to avoid, as far as
reasonably practicable, the risk of accidental
electric shocks to the user, professional or lay, or
other person both during normal use of the device
and in the event of a single fault condition in the
device, provided the IVD medical device is
installed and maintained as indicated by the
manufacturer.
YES
Recognised
standard
EN 61326-2-6:2006
CE-012
YES
Recognised
standard
EN 61326-2-6:2006
CE-012
YES
Recognised
standard
EN 61010-2-101:2002 CE-013
12 C9 Protection against mechanical and thermal risks
C9.1 IVD medical devices should be designed and
manufactured in such a way as to protect the user,
professional or lay, or other person against
mechanical risks connected with, for example,
resistance to movement, instability and moving
parts.
C9.2 Where there are risks due to the presence of moving
parts, risks due to break-up or detachment, or
leakage of substances, then appropriate protection
means must be incorporated.
C9.3 IVD medical devices should be designed and
manufactured in such a way as to reduce to the
lowest practicable level the risks arising from
vibration generated by the devices, taking account
of technical progress and of the means available for
limiting vibrations, particularly at source, unless the
vibrations are part of the specified performance.
C9.4 IVD medical devices should be designed and
manufactured in such a way as to reduce to the
lowest practicable level the risks arising from the
noise emitted, taking account of technical progress
EN ISO 14971:2012
CE-019
YES
Recognised
standard
YES
Recognised
standard
EN 61010-2-101:2002 CE-013
YES
Recognised
standard
EN 61010-2-101:2002 CE-013
YES
Recognised
standard
EN 61010-2-101:2002 CE-013
13 and of the means available to reduce noise,
particularly at source.
C9.5 Terminals and connectors to the electricity, gas or
hydraulic and pneumatic energy supplies which the
user, professional or lay, or other person has to
handle should be designed and constructed in such
a way as to minimize all possible risks.
C9.6 IVD medical devices should be designed and
manufactured in such a way as to reduce to the
lowest practicable level, the risk of error when
certain parts within the device are intended to be
connected or reconnected before or during use.
C9.7 Accessible parts of the IVD medical devices
(excluding the parts or areas intended to supply heat
or reach given temperatures) and their surroundings
should not attain potentially dangerous
temperatures under normal use.
C10 Protection against the risks posed by IVD
medical devices for self-testing
C10.1 IVD medical devices for self-testing should be
designed and manufactured in such a way that they
perform appropriately for their intended purpose
NO。 This device does
not contain these
terminals
connectors.
or
YES
Recognised
standard
EN ISO 14971:2012
CE-019
YES
Recognised
standard
EN 61010-2-101:2002
CE-013
YES
Recognised
standard
EN 62366:2008
EN13612:2002
CE-017
CE-014 性能测试报告
14 taking into account the skills and the means
available to lay persons and the influence resulting
from variation that can reasonably be anticipated in
the lay person’s technique and environment. The
information and instructions provided by the
manufacturer should be easy for the lay person to
understand and apply.
C10.2 IVD medical devices for self-testing should be
designed and manufactured in such a way as to
reduce as far as reasonably practicable the risk of
error by the lay person in the handling of the device
and, if applicable, the specimen, and also in the
interpretation of results.
C10.3 IVD medical devices for self-testing should, where
reasonably possible, include a procedure by which
the lay person can verify that, at the time of use, the
product will perform as intended by the
manufacturer.
C11 Label and Instructions for Use
C11.1 Users should be provided with the information
needed to identify the manufacturer, to use the
device safely and to ensure the intended
performance, taking account of their training and
YES
Recognised
standard
EN 62366:2008
EN13612:2002
CE-017
CE-014
YES
Recognised
standard
EN 62366:2008
EN ISO 15197:2003
CE-017
CE-011
EN ISO 15197:2003
YES
Recognised
standard
CE-011
15 knowledge. This information should be easily
understood.
C12 Performance evaluation including analytical
performance and, where appropriate, clinical
performance
C12.1 For an IVD medical device a performance
evaluation should be conducted in accordance with
GHTF guidance. The performance evaluation
should review analytical performance data and,
where appropriate, clinical performance data in the
form of any:
•
•
•
YES
Recognised
standard
EN13612:2002
CE-014
literature;
performance study reports; and
experience gained by routine diagnostic testing.
to establish that the IVD medical device achieves its
intended performance during normal conditions of use and
that the known, and foreseeable risks, and any undesirable
effects, are minimised and acceptable when weighed
against the benefits of the intended performance.
C12.2 Clinical performance studies using specimens from
human subjects should be carried out in accordance
with the spirit of the Declaration of Helsinki. This
NO Performance
studies
venous
using
blood
samples are come
16 includes every step in the clinical performance
study from first consideration of the need and
justification of the study to publication of the
results. In addition, some countries may have
specific regulatory requirements for informed
consent.
from the hospital;
and the users are
voluntary
evaluation.
to
perform the user
Compiled by:
Reviewed by:
Approved by:
17